12
INSTALLATION
1.
Removing the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent
can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After
the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with
a dry cloth.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2.
Adjusting the leveling foot
Extend the leveling feet by rotating them counterclockwise to contact them to the floor. Ensure the unit
is level. (Fig. 1)
3.
Fixing the unit
Two fixtures are attached to the rear of the frame. Fix the frame to the wall with these fixtures and rope
or chain. (Fig. 2)
4.
Ground (earth)
The ground (earth) is for preventing the electric shock in the case of the electrical insulation is somehow
degraded. Always ground the unit at the time of installation.
WARNING
Use a power supply outlet with ground (earth)
to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
Such
grounding may cause electric shock in the case of an incomplete circuit.
5. Installing the earth leakage breakers
This product is to be connected to a dedicated circuit protected by branch circuit breaker.
Contact our sales representative or agent.
The unit’s power supply cord must be connected to its own dedicated and properly wired
connection outlet where no other plugs are connected.
Power supply cord
Wire size
1.5mm
2
30/0.245
Type CTCE
/
GTVD /
GTSA
VDE
Rating
250 VAC, 16 A
Terminal
Type JST
2-5
Size
Outer diameter; 9.5 mm
Inner diameter; 5.3 mm
Fixture
Fig. 2
Fig.1
Leveling foot