38
INSTALLATION OF TEMPERATURE RECORDER
Attachment of MTR-G04A or MTR-G04C
1.
Remove the right bottom shelf (right bottom drawer for
MPR-1014R) in the chamber.
2.
Unscrew the screw fixing the panel cover at the space
for temperature recorder and open the panel cover. [Fig. 1]
The screw has a cap. Remove the cap beforehand.
3.
Remove the panel cover from the shaft by pushing the
shaft edge to the right and left side respectively. [Fig. 1]
4.
Fix the temperature recorder to the recorder fitting in the
recorder fixing (MPR-S7) referring to the instruction
manual enclosed with the recorder fixing. [Fig. 2]
5.
Unscrew 2 screws on the air intake vent cover (right
side) on the front bottom and then remove the air intake
vent cover. [Fig. 3]
6.
Remove the putty covering the sensor port and take out
the insulation in it. Pass the recorder sensor into the
chamber through the sensor port from the space for
temperature recorder. [Fig. 4]
Screw
Air intake vent cover
Fig. 2
Shaft
Panel cover
Recorder fitting
<overhead view of right chamber>
<overhead view of right chamber>
Fig. 1
Fig. 3
Fig. 4
Sensor port