CHAMBER TEMPERATURE SETTING
Table 1 shows the basic procedure for setting the chamber temperature. Perform key operations in the
sequence indicated in the table. The example in the table is based on the assumption that the desired
temperature is 4
o
C.
Note:
The chamber temperature is set to 5
o
C at the factory.
Table 1 Basic operation sequence (Example: Chamber set temperature 4
o
C)
Description of operation
Key operated
Indication after operation
1 Turn the power switch ON.
----
The current chamber temperature is
displayed.
2 Press set key.
SET
The first digit is flashed.
3
Press numerical value shift key and
scroll the figure to 4.
When pressed, the figure of settable
digit changes.
4 Press set key.
SET
Set temperature is memorized and the
current chamber temperature is
displayed.
Note:
The temperature set mode returns to the temperature display mode automatically when 90 seconds has
passed without any key operation. In this case, the chamber temperature setting is not accepted.
The available set range of chamber temperature is between 2
o
C and 23
o
C. The partial freezing may
occurred if the set temperature is lower than 3
o
C.
KEY LOCK FUNCTION
This unit is provided with a key lock function. When the key lock is ON, change of temperature setting
through the key pad is not available. The key lock is set in OFF at the factory.
Display Mode
Function
L 0
Key lock is OFF
Enable to change of temperature setting
L 1
Key lock is ON
Disable to change of temperature setting
Table 2 Procedure for key lock setting (change from key lock OFF to key lock ON)
Description of operation
Key operated
Indication after operation
1
----
The current chamber temperature is
displayed.
2 Press digit shift key for 5 seconds.
The first digit is flashed.
3
Press numerical value shift key and
scroll the figure to 1.
When pressed, the figure of settable
digit changes.
4 Press set key.
SET
The key lock is set to ON.
The current chamber temperature is
displayed.