- 51 -
E
missions
T
EsT
C
omplianCE
E
lECTromagnETiC
E
nvironmEnT
- g
uidanCE
RF emissions
CISPR 11
Class B
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/Flicker
emissions
IEC 61000-3-3
Complies
Emission of Radio Frequency
Energy
RTCA/DO-160G Section 21
Category M
This device is suitable for use onboard
commercial airplanes inside passenger cabin.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity – This device is
intended for use in the electromagnetic environment specified below. The user of this device
should make sure it is used in such an environment.
i
mmuniTy
T
EsT
iEC 60601 T
EsT
l
EvEl
C
omplianCE
l
EvEl
E
lECTromagnETiC
E
nvironmEnT
-
g
uidanCE
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15 kV air discharges
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15 kV air discharges
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 35%.
Electrical fast Transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines, 100 kHz repetition
rate
±2 kV for power supply
lines, 100 kHz repetition
rate
Mains power quality
should be that of
a typical home or
hospital environment.
±1 kV for input-output
lines; 100 kHz repetition
rate
N/A- Device does not
have user I/O lines that
are longer than 3m in
length.