21
Mode
Setting
S
S/T
T
PC
Flex Type
Bi-Flex
(only if AVAPS Off)
n/a
n/a
n/a
Flex
Off, 1, 2, 3
n/a
n/a
n/a
AVAPS
Off, On
(only if Flex Off)
Off, On
Off, On
Off, On
IPAP
EPAP-30 cm H
2
O
0.5 cm H
2
O res
EPAP-30 cm H
2
O
0.5 cm H
2
O res
EPAP-30 cm H
2
O
0.5 cm H
2
O res
EPAP-30 cm H
2
O
0.5 cm H
2
O res
IPAP Max Pressure
Min IPAP-30 cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
Min IPAP-30 cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
Min IPAP-30 cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
Min IPAP-30 cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
IPAP Min Pressure
4-Max IPAP cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
4-Max IPAP cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
4-Max IPAP cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
4-Max IPAP cm H
2
O
0.5 cm H
2
O res
(AVAPS ON)
EPAP
4-IPAP (25) cm H
2
O
0.5 cm H
2
O res
4-IPAP (25) cm H
2
O
0.5 cm H
2
O res
4-IPAP (25) cm H
2
O
0.5 cm H
2
O res
4-IPAP (25) cm H
2
O
0.5 cm H
2
O res
AVAPS Rate
0.5-5.0 cm H
2
O/min
0.5 res (AVAPS ON)
0.5-5.0 cm H
2
O/min
0.5 res (AVAPS ON)
0.5-5.0 cm H
2
O/min
0.5 res (AVAPS ON)
0.5-5.0 cm H
2
O/min
0.5 res (AVAPS ON)
Tidal Volume
200-1500 ml
10 ml res (AVAPS ON)
200-1500 ml
50 ml res (AVAPS ON)
200-1500 ml
50 ml res (AVAPS ON)
200-1500 ml
50 ml res (AVAPS ON)
Breath Rate
n/a
0-40 BPM
4-40 BPM
0-40 BPM
Inspiratory Time
n/a
0.5-3.0 sec
0.1 sec res
0.5-3.0 sec
0.1 sec res
0.5-3.0 sec
0.1 sec res
Trigger Type
Auto-Trak,
Auto-Trak Sensitive,
Flow Trigger
Auto-Trak,
Auto-Trak Sensitive,
Flow Trigger
n/a
Auto-Trak,
Auto-Trak Sensitive,
Flow Trigger
Flow Trigger
Sensitivity
1.0-9.0 l/min
1 l/min res (only if Flow
Trigger enabled)
1.0-9.0 l/min
1 l/min res (only if Flow
Trigger enabled)
n/a
(only if Flow Trigger
enabled)
Flow Cycle
Sensitivity
10-90%
1% res (only if Flow
Trigger enabled)
10-90%
1% res (only if Flow
Trigger enabled)
n/a
n/a
Rise Time
1-6 (100 - 600 ms)
(only if Flex Off)
1-6 (100 - 600 ms)
1-6 (100 - 600 ms)
1-6 (100 - 600 ms)
Ramp Length
0-45 min
(5 min res)
0-45 min
(5 min res)
0-45 min
(5 min res)
0-45 min
(5 min res)
Ramp Start
Pressure
4-EPAP cm H
2
O
(0.5 cm H
2
O res)
4-EPAP cm H
2
O
(0.5 cm H
2
O res)
4-EPAP cm H
2
O
(0.5 cm H
2
O res)
4-EPAP cm H
2
O
(0.5 cm H
2
O res)
Low Tidal Volume
Alarm
Off, On
(AVAPS ON)
Off, On
(AVAPS ON)
Off, On
(AVAPS ON)
Off, On
(AVAPS ON)
Note: “Res” stands for resolution, which is the incremental value for that setting.