30
English
Emissions
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Complies
Class A Category
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
The device is suitable for use in locations in
residential environments and in
establishments directly connected to a low
voltage power supply network which supplies
buildings used for domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The Philips Avent single/double electric breast pump Premium Plus is intended for use in the
electromagnetic environment listed below. The user of the device should ensure that it is used in such
an environment to prevent improper operation.
Immunity test
Immunity test level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Floors can be a source of static electricity.
When the air is very dry (<20
% RH) the user is
advised to prefer to use the device in a room
with concrete or wooden floors (instead of
using it in a room with a synthetic floor).
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80
% AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
The device is immune against any RF wireless
communication device such as smart phone,
Wi-Fi or Bluetooth device at a distance of
30
cm or more. Avoid a closer distance as this
may result in improper operation.
Rated power frequency
magnetic fields IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV
100 kHz repetition
frequency
Mains power quality expected of a typical
commercial or residential environment.
Surges
IEC 61000-4-5
±1 kV Line-to-line
±2 kV Line-to-ground
Mains power quality expected of a typical
commercial or residential environment.