29
English
English
Immunity test
Immunity test level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Floors can be a source of static electricity.
When the air is very dry (<20
% RH) the user is
advised to prefer to use the device in a room
with concrete or wooden floors (instead of
using it in a room with a synthetic floor).
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80
% AM at 1 kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
385 (27), 450 (28), 710 (9),
745 (9), 780 (9), 810 (28),
870 (28), 930 (28), 1720
(28), 1845 (28), 1970 (28),
2450 (28), 5240 (9), 5500
(9), 5785 (9) MHz (V/m)
The device is immune against any RF wireless
communication device such as smart phone,
Wi-Fi or Bluetooth device at a distance of
30
cm or more. Avoid a closer distance as this
may result in improper operation.
Rated power frequency
magnetic fields IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV
100 kHz repetition
frequency
Mains power quality expected of a typical
commercial or residential environment.
Surges
IEC 61000-4-5
±1 kV Line-to-line
±2 kV Line-to-ground
Mains power quality expected of a typical
commercial or residential environment.
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 Vrms, before
modulation is applied
0.15 MHz – 80 MHz
6 Vrms, before
modulation is applied, in
ISM and amateur radio
bands between 0.15 MHz
and 80 MHz
80
% AM at 1 kHz
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
0 % UT; 0.5 cycle at 0°,
45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle and 70 %
UT; 25/30 cycles single
phase at 0°
0 % UT; 250/300 cycle
Mains power quality expected of a typical
commercial or residential environment. At
power interruptions the breast pump will
automatically switch-over to battery mode
without interruption of the function.
Note: UT is the AC mains voltage prior to application of the test level. If in the unlikely event the Philips
Avent single/double electric breast pump Premium Plus is locked inoperative in a function by
disturbances from other devices beyond the levels stated in the table above, the device will shut-off
completely and restart. In that case make sure the distance to the other device causing this
disturbance is enlarged.