Internal LED Indicators
5-8
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
5.3.2
LED’s on the Detector Power Supply
Five LED’s inside the detector power supply provide information about the integrity of the output
voltages.
The LED’s are visible through the perforations of the top cover.
Figure 5.6 LED’s on the detector power supply
LED COLOR
SIGNIFIES
HVHIGH
RED
High voltage higher than specified setting
HVGOOD
GREEN
High voltage at specified setting
HVLOW
RED
High voltage lower than specified setting or not enabled
LVGOOD
GREEN
All low voltages in specified range
LVLOW
RED
One or more low voltages lower than specified range
Table 5.1 Significance of LED’s on the detector power supply
5.3.3
LED’s on the Detector EDC Boards
An LED is located on the rear panel of the EDC board (see Figure 5.7 on page 5-9). The LED of
the EDC board configured as DET0 (detector 1) should be flashing green at 0.5 Hz indicating that
the EDC board is up running. The LED on the EDC board configured as DET1 (detector 2) should
be flashing red at the same rate. At power up or reset, the LED will be constantly red.
A flashing red light indicates that an error was detected during power up.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......