Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-41
Procedure Details
If one of the nodes is missing, it indicates a
problem.
The three examples on the right (Series III
systems) illustrate the possible locations of
problems.
Between acquisition PC
and FireWire hub
Between FireWire hub
and detector 2 EDC board
Between the two EDC
modules
In CardioMD Series I and II systems, the
FireWire comprises three nodes:
•
The controlling node, i.e. the acquisition
PC, shown as a white box
•
The two regular nodes, i.e. the two
detector EDC modules, shown as green
boxes.
The FireViewer window shown in the example
indicates normal communication.
If one of the two green nodes is missing, it
indicates a problem.
The two examples on the right illustrate the
possible locations of problems.
Between acquisition PC
and detector 2 EDC board
Between the two EDC
modules
5.6.4
Acquisition PC Error Logs
The log file information recorded on the CardioMD acquisition PC is a useful tool for diagnosing
system failures and intermittencies. Technical support personnel may request that copies of these
files be provided in an effort to better diagnose system problem. The information presented in this
section is for reference only. Detailed information regarding the contents of these files is not
provided.
5.6.4.1
System Log Directory
Every time the CardioMD application is started, a new log file is created in the directory
\Cardiocam\SystemLog
. The name of the log file is, for example:
Log{May 28 2004 - 11.04.07}.txt
Thus, each log file has a unique name indicating which log file is most recent. It is also possible to
find old log files related to a specific date and time.
III
I, II
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......