Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-49
Procedure Details
14.
When the Persistence page reappears,
press [F3].
15.
The CardioMD application displays the
study files stored in the folder you have
just set up.
16.
Select either the patient (uppermost list) or
the study/studies to be imported and click
Transfer.
More detailed operating instructions can be
found in the CardioMD Operator’s Manual,
Chapter 4
Operating Instructions
. See the
sections
Exporting a Patient
and
Exporting a
Study
.
17.
When the data transfer is complete, click
Done to return to Persistence page.
The study files have been imported into the
patient database.
When done with the import operation, it is
important to delete the folder node:
18.
Proceed as explained in step 1 – 5 to
display the Nodes Configuration dialog.
19.
Select the Folder entry and click the
Delete button.
20.
When prompted to confirm the delete
operation, click OK.
21.
Click Exit.
22.
Close down and then re-start the
CardioMD application to disable the
service password.
Delete
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......