Repair Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
6-5
Procedure Details
acquisition PC.
6.
Open the latest log file in the acquisition
PC directory
C:\cardiocam\SystemLog
.
7.
Check that both motion controllers have
the same software version.
See Chapter 7
Software Update Procedures
,
the section
Checking Installed Software
Versions
, for detailed instructions.
8.
If they do not, update motion controller
software.
See Chapter 7
Software Update Procedures
for instructions.
6.2.7
Replacing the Power Supply
Procedure Details
To replace the AC mains power supply:
1.
Unplug all cables to the power supply.
2.
Unscrew the four bolts fixating the power
supply to the mounting plate inside the
table console.
Note
. The mounting plate works as heat sink
for the power supply. The power supply must
be firmly mounted on the plate when powered.