Detector Repairs
6-14
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
11.
Remove the detector cover again.
12.
Using a 2 mm Allen key, tighten the upper
set screw on the back of each sensor
support.
13.
Remove the alignment tools and place
them in their original stand-by locations.
14.
Screw the collision sensors back into their
original locations and connect the cables.
15.
Remount the detector cover using all its
screws. Make sure that the M4 Allen
screws have their lock washers applied.
16.
Check the cover collision by rocking the
cover in both directions around its pivot
point.
A collision should be detected in both
directions.
17.
Rotate the detector to 45° and check that
collisions in both directions can be cleared
when using the hand controller Collision
Override button.
18.
Rotate the detector to – 45° and check that
collisions in both directions can be cleared
when using the Collision Override button.