Gantry Software Updates
7-8
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Preparing for Firmware Upgrade
Procedure Details
1.
Remove the detector top cover to access
the EDC modules. See instructions in
Chapter 6
Repair Procedures
.
2.
Remove the ECG cable from the ECG port
labeled A1 RS232.
3.
Connect one end of the serial interface
cable to this port.
If you are using a laptop acquisition PC:
4.
Use the USB to RS 232 adapter to connect
the opposite end of the serial interface ca-
ble to one of the acquisition PC’s USB
ports.
See section 7.2.2, page 7-2 for instructions on
determining which COM port you are using.
If you are using a tower acquisition PC:
5.
Connect the opposite end of the cable to
the acquisition PC’s COM1 serial port.
6.
Insert the software installation CD-ROM
in the acquisition PC’s CD-ROM drive.
Upgrading Micro Controller Firmware
Procedure Details
To load micro controller firmware to the EDC
Board, you must start the Flash program. Pro-
ceed as follows:
1.
In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-B02 or
later
on the CD-ROM.
2.
In this folder, double-click the icon
flash.exe
to start the Flash program.
III
I, II
Serial interface
cable
ECG cable
III
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......