Shipping, Handling, Installation and Storage
8-14
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.2.2.1
Equipment Moved in Boxes on Pallet Jacks
Door width
62 inch (150 cm)
Corridor width (90° angle)
67 inch (175 cm)
Door height
80 inch (200 cm)
Elevator width (if required)
62 inch (150 cm)
Elevator depth (if required)
55 inch (140 cm)
Elevator height (if required)
80 inch (200 cm)
Table 8.4 Passage clearances for moving equiment in boxes on pallet jacks
Figure 8.8 Minimum clearance for turns
8.2.2.2
Equipment Moved on Dolly Shipped with the System
Door width (with/without jack beams mounted)
36 inch (91 cm) / 30 inch (75 cm)
Corridor width (90° angle)
42 inch (105 cm)
Door height
80 inch (200 cm)
Elevator width (if required) (with/without jack beams mounted) 36 inch (91 cm) / 30 inch (75 cm)
Elevator depth (if required)
60 inch (150 cm)
Elevator height (if required)
80 inch (200 cm)
Table 8.5 Passage clearances for moving equipment on dolly