Installation
8-22
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.
Position the gantry and lower the jacks. The gantry will be guided into the plates for the last
12 mm. Make sure that the gantry can be mounted to the plates with 4 pcs. M10 screws (2 in
each end bar).
9.
Insert the neoprene weather strip between the outside edge of the gantry frame and the floor,
to prevent dirt from entering underneath gantry.
10.
Using the dolly wheels, build trolley to transport the table base to the installation room..
11.
Using the jack beams mounted to the table base, remove the pallet and lower table base on top
of trolley. Remove the jack beams, if needed, for narrow doors.
12.
Transport the table base to the installation room. Place it close to its final position. If the jack
beams were removed, remount them, jack up and move the trolley away.
13.
Lower the table base and remove the front jack beam leaving the rear end on the dolly beam
about 25 mm (1 inch) above the floor. Place one of the protective plastic feet under the front
end (close to the gantry) and mount 2 long bolts through the gantry frame as guide.
14.
Push the table base in position. The table base is guided the last 5 mm by a key. Remove the
long bolts and mount the table base to the gantry with 2 screws, one at the time. Tighten the
screws and then loosen them by half a turn. Lower the rear end and remove the jack beam.
15.
Mount the last 2 M12
×
30 anchor screws and washers in the rear end of table base and
tighten. Tighten the 4 anchor screws in the mounting plates for the gantry.
16.
Remove 4 mounting screws in the gantry (2 in each end bar) Using the special tool on the
leveling screws and a spirit level, align the gantry to be level in horizontal plane. Remount the
4 screws and tighten.
17.
Remove the 2 screws in the middle of the side bar and adjust the leveling screws. Remount the
2 screws and tighten.
18.
Remove the 2 mounting screws in the table base rear end and align the cradle to be in level.
Remount the 2 screws in the rear end.
19.
Tighten the 2 screws holding the gantry and table base together. Remove the 2 shipping locks
for the x-axis.
20.
Mount the table support to the left sidebar of the gantry base.
21.
Connect the 3 cables from the gantry to the table base.
22.
Assemble and mount the acquisition PC option and cable it to the gantry. Connect the hand
controller.
23.
Configure input voltage selection; connect power and power up the system.
24.
Tighten rear anchor bolts. Using the hand controller, move all axes. Leave the detector in
upper position.
25.
Remove the detector insulation (after 24 h of constant temperature).
26.
Mount patient table pallet and mattress.
27.
Check motion limits.
28.
Mount stays for base cover.
29.
Mount covers.
30.
Set up and connect ECG gate.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......