Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-31
Procedure Details
1.
Using the template shipped with the
system and a plan of the room layout,
measure the minimum distance to the
walls.
2.
Position the template.
3.
Use adhesive tape to fix the template to
the floor.
8.3.3.5
Drilling Holes and Setting Anchors
Procedure Details
1.
Drill 6 holes, 2¼” (55 mm) deep, for
anchors through the template. Use the
15 mm masonry drill.
You may use a hammer drill for a pilot hole.
2.
Remove the template and clean the holes.
Note
. If rebars or tensioning cables in the floor obstruct drilling, use the alternative holes
provided.
Procedure Details
3.
Use a mandrel to set the anchors.
Summary of Contents for CARDIOMD