Installation
8-32
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.3.3.6
Transporting the Gantry to the Installation Room
Note
. If the corridor has a 90° corner, the corridor width must be at least 42” (105 cm). See Figure
8.18. The gantry on its dolly with the jack beams mounted can go through 91 cm (36”) doorways.
Without the jack beams, the gantry can pass a doorway of 75 cm (30”).
Figure 8.18 Passage clearance for gantry
Procedure Details
1.
Roll the gantry in position over the
anchors holes.
2.
If the jack beams have been removed,
remount these now.
3.
Jack up and dismount the dolly wheels.