Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-75
8.3.4
Software Setup
This section provides instructions for setting up software after installing the CardioMD system
and laptop acquisition PC. On the laptop acquisition PC, the acquisition software runs under
Microsoft Windows XP. However, this section also includes some instructions on software setup
under Windows 2000 in case you need to perform these tasks on a tower acquisition PC
(CardioMD Series I and II) running Windows 2000.
8.3.4.1
Local Area Network Setup
This section explains how to set up the local area network. This involves actions that may
jeopardize the Windows network operation. Extreme caution must be taken when modifying the
network configuration.
To perform network setup, you need administrator privileges on the acquisition PC. If your
acquisition PC is running Windows XP, you automatically get administrator privileges. For
Windows 2000, proceed as follows:
Procedure Details
If the CardioMD acquisition is running:
1.
Ensure that the acquisition PC is
displaying the Persistence page. Click the
Close button in the upper right corner of
the window to close the CardioMD
application.
2.
Click the Windows Start button.
3.
Select
Shut Down
4.
In the drop-down list, select
Log off
CardioMD
.
If the menu item says
Log off Administrator
,
click Cancel and proceed to the section
Setting
Up Acquisition PC IP Addresses
below.
5.
Click OK.
The Log on to Windows dialog appears.
6.
In the User name field, enter
Administrator
. Ensure that the
Password field is left blank.
7.
Click OK.
Setting Up Acquisition PC IP Addresses
Contact the institution’s systems administrator for setup parameters for the local area network.
The network or systems administrator must provide the following information:
•
An IP address for the CardioMD system
•
A network node name for the CardioMD
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......