Installation
8-84
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
The contents of the
institution
file look
similar to the example shown to the right.
8.
Edit the institution name, address and ID.
9.
Select
File
,
Save
to save the modified file.
10.
Select
File
,
Exit
to close down the
Notepad program.
11.
Double-click the CardioMD icon on the
desktop to start the CardioMD application.
12.
Check that the correct institution name and
ID appear during startup.
8.3.4.5
Setting Up Units for Patient Height and Weight
Procedure Details
1.
On the acquisition PC, ensure that the
CardioMD application is displaying the
Persistence page.
2.
Click the Close button in the upper right
corner of the application window to close
down the application.
3.
In the Windows Start menu, select
Programs
,
Accessories
,
Notepad
.
The Windows Notepad application starts.
4.
In the Notepad window, select
File
,
Open
.