Detailed Description
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
3-7
Connector
Name
Voltage
Function
OC1
115V AC
Series III: ACQUISITION PC and ECG gate
Series I and II: ACQUISITION PC
C1 55V
DC
SPARE
C2
55V DC
DETECTOR 1 EDC BOARD AND DETECTOR PSU
C3
55V DC
DETECTOR 2 EDC BOARD
C4 55V
DC
BOTH
MOTION CONTROLLERS
ROTATION AND VERTICAL AMPLIFIER*
C7 24V
DC
SAFETY
BOARD
* if these are of the type that requires 55 V DC input
Table 3.1 Continuous outputs
Connector Name
Voltage
Function
OC2
180V DC
ROTATE DRIVE AMPLIFIER
OC3
180V DC
Y DRIVE AMPLIFIER
C5
55V DC
X DRIVE AMPLIFIER
C6
24V DC
TABLE DRIVE
Table 3.2 Safety interrupted outputs
Connector C8 goes to the Safety Stop circuit. This input is used to interrupt power to the motors in
case of collision, motion controller software fault, or if one of the E-Stop buttons is pressed.
The electronics within CardioMD, including the detectors, are operated on uninterruptible 55 V
DC. Local DC-DC converters internal to the electronic modules convert the 55 V DC supply
voltage to the voltage levels used internally (i.e., 3.3 V DC, +/– 5 V DC, 12 V DC, +/– 15 V DC
and HV).
The AC power supply contains a number of fuses. The access and location of these fuses are
indicated in Figure 3.6 on page 3-6. The electrical location of the fuses is shown in Figure 3.7
below.
MAINS fuses
180 V POWER
CORRECTION
55 V POWER
CORRECTION
115V
32 V
115V
to PC
SECONDARY FUSES
ISOLATING TRANSFORMER
Figure 3.7 Electrical location of fuses in power supply
III
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......