NEMA Energy Resolution
10-6
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
sult.
18.
Replace the NEMA Mask with the colli-
mator.
19.
Click Exit to return to the NEMA menu.
Note
. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.