NEMA Performance Measurement
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
10-11
Procedure Details
When both the X and Y measurements have
been completed, a pop-up similar to the one
shown on the right appears.
14.
Note the results and click OK to close the
pop-up.
Resolution specifications are:
UFOV FWHM
≤
3.7
mm
CFOV FWHM
≤
3.7
mm
UFOV FWTM
≤
7.6
mm
CFOV FWTM
≤
7.6
mm
Linearity specifications are:
UFOV absolute
≤
0.5
mm
CFOV absolute
≤
0.5
mm
UFOV differential
≤
0.2
mm
CFOV differential
≤
0.2
mm.
15.
Replace the NEMA Y line phantom with
the collimator.
16.
Click Exit to return to the NEMA menu.
Note
. To measure the other detector, place the CardioMD system in the patient load position be-
fore repeating the procedure outlined above.