Planned Maintenance
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
11-5
Procedure Details
2.
Use the grease gun to apply lithium grease
through the zerk fitting on the ball screw
nut as shown in the photo on the right.
3.
Remove the rotate gear cover and use the
grease gun to apply lithium grease through
the two zerk fittings on the bearing shown
in the photo on the right.
11.3.5
Deleting Old Log Files
The CardioMD stores information logging the operation of the system in log files in the directory
C:\cardiocam\SystemLog
. If the system log directory is not cleaned up from time to time, the
acquisition PC’s harddisk will eventually be filled up with log files. Therefore, on every service
visit, you should delete all log files that are more than one month old.