Repair Procedures
12-44
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.5.3.3
Replacing the AC Source Motion Board
Procedure Details
Before starting to replace the AC source
motion board:
1.
Dismount the transmission scanner from
the detector (see section 12.5.1.1, page 12-
29).
2.
Detach the line source compartment from
the bracket (see section 12.5.1.2, page 12-
33).
3.
Remove line source compartment covers
(see page 12-41).
4.
Detach the connecting cable (see section
12.5.3.2, page 12-43).
When all of the above steps have been
performed, remove the collision detection
shutter end cover:
5.
Use a T08 Torx key to remove the 3
screws indicated in the photo on the right.
6.
Remove the cover.