Repair Procedures
12-58
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
8.
Use a 2.5 mm Allen key to remove the
four screws holding the shutter indicator /
manual lever to the slide bearing.
Note
. Mind the spring holding the actuator
bracket.
9.
Remove the shutter indicator / manual
lever.