CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-85
Procedure Details
16.
With one hand, press the actuator bracket
upwards to open the shutter.
17.
Use the transportation screws to lock the
line source shutter in the Open position.
When the line source shutter is locked open,
the acquisition PC issues a pop-up warning
and emits a continuous audible warning signal.
This is normal.
WARNING
When the line source shutter is open, a collimated beam of radiation is emitted
from the line source. Make sure to
never
leave the system unattended while the
transportation screws are locking the line source shutter in the open position.
CAUTION
RADIOACTIVE
MATERIAL