Installation
12-98
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
12.
Plug the orange-colored 3-way plug from
the AC Option power supply into
connector C6 (24V EM-ST) on the main
PSU.
13.
Plug the white 3-way plug from the AC
Option power supply into connector OC1
on the Main PSU (115VAC) and attach
the associated ground wire with the M4
screw.
14.
Unplug the cable connector C4 from the
main PSU connector panel.
15.
Connect the 55 V split cable 9CBL1482 to
the main PSU connector panel by inserting
the plugs C2 and C4 into the connectors
labeled C2 and C4, respectively.
16.
Connect the opposite end of the 55 V split
cable (labeled C4) to the cable connector
C4 that was removed from the main PSU
connector panel in step 14.
C4 connector
C2 connector
C4 cable
connector