Calibration
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
4-11
4.2.3.3
PMT Calibration Procedure
To be performed:
At installation, and when the EDC board, PMT tube, X-E
board, Y board or crystal is replaced.
Detector:
One detector at the time, without collimator.
Source:
Tc-99m or Co-57. 40 MBq or 1 mCi in the source holder.
Total calibration time:
Approximately 10 minutes per detector.
Calibration file created:
C:/Cardiocam/calibrations/pmtcal.det1
and
det2
respectively.
Note
. If the CardioMD system is equipped with the AC Option, the transmission scanners must be
removed before PMT calibration can be performed. See Chapter 12
CardioMD AC Option
for
instructions.
Note
. The detector cover must be in place and the detector electronics must be at normal operating
temperature before this procedure is started.
Procedure Details
1.
Ensure that the CardioMD system is in the
patient load position.
2.
On the Persistence page, click the Tools
button.
3.
When the Calibration & Configuration
menu appears, click Service.
4.
If prompted, enter the service password
(see page 4-9 for instructions).
5.
In the Service Calibrations menu, click
PMT.
The PMT page appears.
6.
Unless a major repair requires a tuning
from scratch, select Update PMT table.
7.
Select the appropriate isotope (either Tc-
99m or Co-57).
8.
Select detector 1 or detector 2 to be cali-
brated.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......