Detector Calibration
4-18
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
3.
On the Persistence page, click the Tools
button.
4.
When the Calibration & Configuration
menu appears, click Service.
5.
If prompted, enter the service password
(see page 4-9 for instructions).
6.
In the Service Calibrations menu, click
GeoCal.
On the GeoCal page:
7.
Depending on the isotope being used, fill
in the High Energy field as follows:
Ba-133: 155
Tl-201: 167
Ga-67: 185.
8.
Fill in the Low Energy field as follows:
Ba-133: 81
Tl-201: 75
Ga-67: 92.
9.
Select the detector to be calibrated.
10.
Click Continue.
11.
Press and hold the hand controller Colli-
sion Override button for a minimum of 4
seconds before proceeding as follows:
12.
While still pressing Collision Override,
click OK to accept automated motions
bringing the system into position for cali-
bration of the selected detector.
Note
. The hand controller Collision Override
button must be kept depressed during all
automated motions to enable power to the
motors.
Note
. Automated motions must be started
from the patient load position.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......