background image

 

EN

       11

 

account, which causes an error in the synchronization of at least 
several seconds. 

2) Establish a Bluetooth link with the DTI-2 and start up the 

DTI-2_BT_StreamingDataViewer program that can be found on the 
internal data storage of the DTI-2. After opening the communication 
in this program synchronization will take place automatically. 

Battery charging 

Battery recharging is done with the supplied USB charger via the USB 
cable, which needs to be inserted into the USB connector that can be 
found behind the fastening hinge. A red LED placed close to the USB 
connector indicates active charging. It takes about 2 hours to charge the 
device from empty to full. The red LED may not switch off when a full 
charge state is realized. The sure way to check whether the battery has a 
full charge is to unplug the USB cable, and to plug it in again: when the 
red LED stays out the battery is fully charged. 

Data Readout 

1) Connect it to a Windows7

/Vista/XP™ PC with the USB cable. A blue 

LED close to the USB connector indicates a functional USB 
connection. 

2) The device mounts as 

‘’Device with Removable Storage’’ on your PC. 

Open the explorer and click the drive associated with the DTI-2 
(should this fail to happen unplug the USB cable and reconnect it). 
When asked to scan and fix the device, please comply, and tick both 
check disk options. After the scan the DTI-2 drive mounts normally. 

3) Copy the data files to a location of your choice on your own computer. 

4) Undock the device after this operation 

Data format and data visualization 

The measured data is stored in a binary file named DTISamples0XXX.dti, 
where XXX is the serial number of the device. This file can be converted 
into a text file with the DTI-2_SampleFileParser-P program which is 
present on the DTI-2 data storage disk. Run this program, select the .dti 
file, and choose Convert DTI Sample File. This will generate a text file 
with all data in lines with this format: 

yyyymmddHHMMSSmmm|NNNN|ev|ttttt|sksks|xxxxx|yyyyy|zzzzz|vb|etete|eieie|rrrrr 

A minus sign may precede the accelerometer data xxxxx, yyyyy or zzzzz. The 
description of the line coding is given in the next Table.

 

Summary of Contents for DTI-2

Page 1: ... 2013 Koninklijke Philips Electronics N V All rights reserved DTI_2_UM_V2 0 8122 357 24302 ...

Page 2: ...2 EN DTI 2 __________________________ __________________________ EN User manual NL Gebruiksaanwijzing DE Benutzerhandbuch FR Mode d emploi SV Användarhandbok SR Korisničko uputstvo ...

Page 3: ...s in the box 6 Overview of the device 6 User interface 7 3 Intended use 7 4 How to use the device 8 Wearing instructions Switching the device on off 8 Bluetooth wireless data transmission 9 Event marker 9 Status button 9 Time synchronization 9 Battery charging 10 Data readout 10 Data format and visualization 10 5 Product information 12 ...

Page 4: ...dropped 9 Battery usage CAUTION To prevent battery leakage which may result in bodily injury property damage or damage to the unit the batteries battery pack or batteries installed shall not be exposed to excessive heat such as sunshine fire or the like Notice Any changes or modifications made to this device that are not expressively approved by Philips may void the user s authority to operate the...

Page 5: ...sed in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or mo...

Page 6: ...helps prevent potentially negative consequences on the environment and human health Your product contains batteries covered by the European Directive 2006 66 EC which cannot be disposed of with normal household waste Please inform yourself about the local rules on separate collection of batteries The correct disposal of batteries helps prevent potentially negative consequences on the environment a...

Page 7: ...rger 1 x quick start guide Overview of the device 1 User interface 2 Fastening hinge the USB connector is behind it 3 Skin conductance electrode 4 Wrist strap 5 Skin temperature sensor 6 Personal fit segments XXS 5 segments L 9 segments XS 6 segments XL 10 segments S 7 segments XXL 11 segments M 8 segments 1 2 3 5 6 4 ...

Page 8: ...nds When the Bluetooth active a blue LED is visible above the Status button 3 Intended use The device is to be used to monitor at the left or right wrist of humans the skin conductance skin temperature ambient temperature ambient light level and 3 dimensional acceleration The sensor data can be streamed live to a receiving station via a Bluetooth wireless link The device is to not to be used as a ...

Page 9: ...Optimal is the loosest strap position without dangling Place the rubber strap on the underside of the wrist such that the skin conductance electrodes contact a desired area of the skin recommended is to center the electrode furthest from the thumb and wrap the strap towards the electronics module Recommended position of the wristband sensor skin conductance electrode for best results Switching the...

Page 10: ...ht button which is the status button The left array shows the skin temperature all LEDs on means 38 degrees Celsius The decrement is 1 degree per LED so one LED lit means 31 degrees Celsius The right array shows the skin conductance normalized to the average of the last minute 3LEDs on Per increment of 1 a LED switches on or off This allows a visualization of the recent skin conductance changes Ti...

Page 11: ...SB cable A blue LED close to the USB connector indicates a functional USB connection 2 The device mounts as Device with Removable Storage on your PC Open the explorer and click the drive associated with the DTI 2 should this fail to happen unplug the USB cable and reconnect it When asked to scan and fix the device please comply and tick both check disk options After the scan the DTI 2 drive mounts...

Page 12: ...5 Counts rrrrr Skin conductance ADC 5 16bit raw ADC output snr Dataset number 8 The visualization can be done using the DTI 2 Data Viewer program This can be installed from the integrated data storage of the DTI 2 Connect the device to a computer with a USB cable The DTI 2 mounts as Device with Removable Storage on your computer Open the explorer and click the drive associated with the DTI 2 shoul...

Page 13: ...icro Siemens range 0 001 micro Siemens sensitivity Silver coated Eyelet A3822C2 Micron electrodes Skin temperature sensor 20 85o C range 0 02o C sensitivity 100o field of view Ambient temperature sensor 40 125o C range 0 5 2o C accuracy best at 40o C Ambient light sensor 24 4 counts µW cm2 responsivity at 640nm for the VIS IR diode 6 9 counts µW cm2 responsivity at 940 nm for the IR diode 3D accel...

Reviews: