Everflo Q
Operating instructions
Warning: Do not use extension cords or electrical adapters.
1. Select a location that allows the machine to draw in room air
without being restricted. Make sure the device is at least 15 to
30cm away from walls, furniture, and especially curtains that could
impede adequate airflow to the device. Keep away from heat.
2.
Plug the power cord into an electrical outlet.
3.
Do either Step a or Step b below
a.
If you are not using a humidifier, connect your nasal cannula to
the Oxygen Outlet Port.
b.
If you are using a humidifier, follow the steps below:
i.
Open the filter door on the back of the device.
ii.
Remove the humidifier connector tube from the back of
the door and replace the filter door.
iii.
Fill your humidifier bottle according to the manufacturer's
instructions.
iv.
Mount the filled humidifier on the top of the EverFlo
device inside the velcro strap, as shown in the illustration
on the right.
v.
Tighten the velcro strap around the bottle and secure it so
it is held firmly in place.
vi.
Connect the humidifier connector tube (that you retrieved
from the filter door) to the Oxygen Outlet Port.
vii.
Connect the other end of the humidifier connector tube to
the top of the humidifier with the elbow in the tubing
facing the front.