36
About Getting started Daily use Warnings
More info
Symptom
Possible causes
Solutions
The transmitter’s
LED light remains
turned OFF when the
transmitter is placed in
the charger
The charger is not turned ON
Verify that the charger’s power plug is correctly connected
The transmitter or charger’s temperature is either too warm
or too cold
Move the charger and transmitter to a location with a temperature b5°C and
+40°C (+41°F and +104°F)
Charging is incomplete. The room temperature exceeds
+35°C (+95°F), which prolongs the charging time. The
charger
has stopped charging to protect the battery.
Reinsert the transmitter into the charger. This completes the charging within
approximately
15 minutes.
The transmitter is incorrectly seated in the charger
Check the charging ports for foreign objects
The transmitter’s LED
light blinks GREEN when
the transmitter is placed
in the charger
The transmitter has not been in use for a prolonged period
of time
Depending on how depleted the battery in the transmitter is, the transmitter
automatically resumes normal charging mode after a timeframe of up to 10 minutes.
Ensure you leave the transmitter in the charger during this process.
The transmitter’s LED
light blinks ORANGE
when the transmitter is
placed in the charger
System error
Contact your hearing care professional
If none of the above solutions work, consult your hearing care professional.
Troubleshooting
More info