M3500B/M5500B Service Manual
4
-
59
Bottom Case Assemblies
4
Figure 4-31 Battery PCA connections
Replacement
1. Slide the wires on the new PCA into the hole in the case.
2. Install the new gasket into the hole. The gasket is fully seated when it
protrudes out evenly around the hole on both the inside and the outside of
the case.
3. Install the Battery Plate Spring onto the standoff on the Battery PCA.
Install the Battery PCA into its slot in the case.
4. Flip the case over and plug in the 3 connectors (2 on Power PCA, 1 in line
to Power Connector).
Gasket
4 Pin
Connector
to Power
PCA
2 Pin connector
to Power PCA
Power PCA
In-line
Connector
to Power
Connector