Bottom Case Assemblies
4
-
62
Removal and Replacement
Power PCA
The Power PCA is located in the bottom case.
Removal
1. Disconnect and remove the defibrillator capacitor. See "Defibrillator
Capacitor" on page 4-60.
2. Disconnect the 3 connectors from the Patient Connector to the Power
PCA at the Power PCA - 2 are High Voltage spade connectors (note the
correct locations before removing) and 1 is a 9 pin connector with a latch.
See Figure 4-33.
3. Disconnect the 2 connectors from the Battery PCA to the Power PCA at
the Power PCA - 1 large 4-pin connector that carries the battery leads (2
red + 2 black wires), and 1 small 2-pin connector with a latch.
4. Disconnect 1 4-pin connector from the Power Connector to the Power
PCA at the Power PCA.
5. Disconnect the large ribbon cable from the Power PCA to the Control
PCA at the Power PCA end. Unlatch and remove the metal bail (do not
bend it).
6. Remove 6 T10 size screws from the Power PCA.
7. Pull the Power PCA up and guide it around the Patient Connector.