Bottom Case Assemblies
4
-
64
Removal and Replacement
Power Connector
Removal
1. Disconnect 1 connector from the Power PCA (4 conductor with latch).
2. Disconnect 1 connector in line to the battery PCA (2 conductor with
latch).
3. Unscrew the large nut on the back of the connector.
4. Pull the connector and its wires and O-ring out of the case.
Replacement
1. Slide the wires of the new connector through the new O-ring that came
with the connector, then insert them through the hole in the case, and then
through the large nut. Use the existing nut.
2. Seat the O-ring carefully on the connector.
3. Align the flat section on the connector with the flat portion of the case
hole.
4. Push the connector into the hole, seating the O-ring into the case.
5. Tighten the large nut snugly but not excessively - the goal is to compress
the O-ring, not crush it hard against the case.
Figure 4-34 Power Connector and Patient Connector
High Voltage Spade
Connectors (2)
9 Pin Connector
Patient Connector
Power Connector
(hidden)