M3500B/M5500B Service Manual
4
-
67
The Language Support Tool
4
The Language Support Tool
The Language Support Tool allows field service personnel to perform several
tasks: 1) to set the language of the Control PCA of the defibrillator; 2) to set
the default display contrast, 3) to enable the SpO
2
option, and 4) to program in
the serial number. These tasks need to be performed under the following cir-
cumstances:
z
The Control PCA has been replaced.
The Control PCA contains all the operating software, configured
for the installed hardware. It also contains the unit’s serial number,
which was programmed in during manufacturing.
The new Control PCA must be programmed to recognize the
hardware installed in this unit, to contain that unit’s
serial number, and to set the lanugage of the unit.
z
The unit has received an upgrade adding the SpO
2
capability.
The added hardware will
not
be automatically recognized. The
Control PCA must be programmed to recognize the new hardware
installed.
z
The unit’s software is being updated.
Software upgrades are periodically released to resolve customer
issues. The Language Support tool enables you to load new software
onto the unit.
Using the Support Tool
1. Have unit power off, and have either a fresh battery installed or a Power
Module connected.
2. Insert the Support Tool data card into the unit.
3. Turn unit power on.
4. Follow the screen prompts.
a.
Press
Yes
to continue.
The message "Upgrade software found Proceed to Program?" is dis-
played.
b.
Press
Yes
to set the unit’s language. Do
not
press any keys or touch the
unit until the process is complete.
c.
Set the default display contrast level. This value is only used when
the configuration is lost. During the normal use, the unit powers up
with the contrast level it was last set to.
d.
Select whether SpO
2
hardware is installed or not as appropriate.
e.
Program in the unit’s serial number
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...