M3500B/M5500B Service Manual
7
-
3
Specifications
7
AED Mode
AED Energy Protocol:
Fixed Energy (150 Joules).
AED Shock Series:
2, 3, or 4.
Shock Series Timer:
off, 30, 60, 90, 120, 150, 180, or 210 seconds.
Text and Voice Prompts:
Extensive text and audible messages guide user
through protocol.
Protocol:
Follows pre-configured settings. Can be configured to meet Ameri-
can Heart Association and European Resuscitation Council guidelines.
AED Controls:
On, Off, Pause/Resume, Analyze/Stop Analysis, Shock, Lead
Select, SpO
2
, HR Alarms, Display Contrast, ECG Gain, Volume, Print Strip,
Print Summary, Annotated Mark Event, Manual Mode Entry.
Indicators:
LCD Display for ECG waveform and text prompts, Audio Alerts,
Voice Prompts, QRS Beeper, Charging Tone, Charge Done Tone, Printer.
Armed Indicators:
Charge Done Tone, Available Energy indicated on dis-
play, Voice Message.
Patient Analysis:
Per protocol, evaluates patient ECG and signal quality to
determine if a shock is appropriate and evaluates connection impedance for
proper defibrillation pad contact.
Shockable Rhythms:
Ventricular fibrillation with amplitude > 100 uV and
wide complex ventricular tachycardia with rates greater than 150 bpm.
Sensitivity and Specificity:
Meets AAMI guidelines.
ECG Monitoring
Inputs:
Single channel ECG may be viewed on display and printed. PADS
ECG is obtained through two multifunction defibrillation electrode pads.
Lead I, II, or III is obtained through the 3-lead ECG cable and separate moni-
toring electrodes. With a 5-lead cable, lead aVR, aVL, aVF, or V can also be
obtained.
Lead Fault:
LEADS OFF message and dashed line appear on the monitor if a
lead electrode or wire becomes disconnected.
Pad Fault:
PADS OFF message and dashed line appear on the monitor if a
pad becomes disconnected.
Heart Rate Display:
Digital readout on display from 15 to 300 bpm, with an
accuracy of 10%.
Heart Rate Alarms:
Configurable pairs of low and high heart rate alarm lim-
its: 30 to 100, 60 to 140, 90 to 160, and 120 to 200 bpm.
Defibrillation Patient Cable Length:
7 ft. (2.13 m).
ECG Cable Length:
10 ft. (3.05 m).
Common Mode Rejection:
Greater than 90 dB measured per AAMI stan-
dards for cardiac monitors (EC 13).
+
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...