Specifications
7
-
6
Specifications
Noninvasive Pacing
Waveform:
Monophasic Truncated Exponential.
Current Pulse Amplitude:
10 mA to 200 mA (5 mA increments); accuracy
10 mA - 50 mA 5 mA, 50 mA - 200 mA 10%.
Pulse Width:
20 ms with ac0, -5 ms.
Rate:
30 ppm to 180 ppm (10 ppm increments); accuracy 1.5%.
Modes:
Demand or Fixed Rate.
Refractory Period:
340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm).
SpO
2
/Pulse Oximetry
Accuracy with:
M1191A sensor -
1 standard deviation 70% to 100%, 2.5%.
M1192A sensor -
1 standard deviation 70% to 100%, 2.5%.
M1194A sensor -
1 standard deviation 70% to 100%, 4.0%.
NELLCOR sensors -
1 standard deviation 80% to 100% 3.0%.
Resolution:
1%.
SpO
2
Alarm Limits:
Three preset low alarm limits: 90, 85, and 80%, default
is off.
INOP Alerts:
Triggered by disconnected sensor, noisy signal, light interfer-
ence or low signal (non-pulsatile).
Event Storage
Internal Event Summary:
The internal Event Summary stores up to 300 events and up to 50 waveforms.
Events can be marked with a Mark Event symbol and, if configured for drug
annotation, the following labels can be added: Epinephrine, Atropine,
Lidocaine, or Other.
The Print Summary key on the front panel is used to print the internal Event
Summary.
Data Card Event Summary:
The Data Card stores continuous ECG waveforms and events on a Type II
PCMCIA card, SanDisk SDP3B 8MB ATA FlashDisk. One 8MB card can
store approximately 2 hours of continuous ECG waveforms and events.
+
+
+
+
+
+
+
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...