Mandatory Testing
2
-
2
Performance Verification and Safety Tests
z
Labels
z
AC or DC Power module
z
Consumables (ECG monitoring electrodes, multifunction defibrilla-
tion pads, printer paper)
z
Main fuse (on Battery PCA)
z
Manual Door Assembly
z
Data Card Door Assembly
z
Battery Eject Assembly
The following testing is required after an
External Repair
or when the out-
come of the service is
No Trouble Found
(when the case has not been
opened)
:
z
Perform the Visual Inspection (page 2-12).
z
Run the Extended Self Test (page 2-17).
z
Print and Verify the System Log (page 2-16).
Printer
If the printer was replaced,
and the case was not opened,
the following tests
are required:
z
Perform the Visual Inspection (page 2-12).
z
Run the Extended Self Test (page 2-17).
z
Run the Printer Test (page 2-21).
z
Print and Verify the System Log (page 2-16).
Internal Repairs
If the case was opened (regardless of what the repair involved), all of the Per-
formance Verification Tests must be performed, beginning with "Visual
Inspection" on page 2-12.