M3500B/M5500B Service Manual
2
-
15
Performance Verification
2
Sp0
2
Functional Check
This check only needs to be performed if SpO
2
is installed.
1. Attach the SpO
2
transducer to your finger.
2. Activate Manual Mode and press the SpO
2
softkey to turn SpO
2
on.
3. The SpO
2
value displayed should be in the range of 95-100%. If the value
is less than 95%, check that your finger is fully inserted into the sensor
and properly positioned.