M3500B/M5500B Service Manual
2
-
17
Diagnostic Tests
2
Extended Self Test
The Extended Self Test checks that all internal processors are operating and
communicating with each other. The instructions below describe how to run
the Extended Self Test.
If all results are as described, the unit passes this test. Return to the main
Diagnostic Test menu by pressing
.
If there is any failure, begin troubleshooting and repairing the unit as needed.
See "Troubleshooting" on page 3-1, and Table 3-6 "Extended Self Test Fail-
ures" on page 3-18.
NOTE
Make sure that no one touches the unit during the self test as that can generate
an erroneous
Not Tested
or
Fail
message.
1. Connect the test load to the pads cable.
2. Access the Diagnostic Test menu as described in "Diagnostic Mode"
above.
3. Select
Extended Self Test
and press
. The printout should appear
similar to Figure 2-3.
Figure 2-3 Sample Extended Self Test Printout
4. Check the printout to verify that the time and date are correct. If they are
not, reset them using the Configuration Mode. See "Configuration Mode"
on page 2-10.
5. The results of the following tests will appear on the display and on the
printout:
z
ROM
Tests the Read Only Memory (ROM).
z
RAM
Tests the Random Access Memory (RAM).
MAIN
ENTER
EXTENDED SELF TEST
18:48 1/19/2000
ROM:
Pass
RAM:
Pass
System:
Pass
Data Card:
Not Tested
CODEC:
Pass
IRDA:
Not Tested
Timebase:
Pass
Defib:
Pass
FE:
Pass
SpO2:
Pass
Pacer:
Pass
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...