M3500B/M5500B Service Manual
2
-
29
Diagnostic Tests
2
Defibrillator Disarm Test
These instructions describe how to test the disarm function.
If all results are as described, the unit passes the test. Return to the main Diag-
nostic Test menu by pressing
. If there is any failure, begin trouble-
shooting and repairing the unit as needed. See "Troubleshooting" on
page 3-1.
1. Connect the defibrillator analyzer to the pads cable. If needed, reset the
analyzer’s display to read 0.
2. From the Diagnostic Menu, Select the
Defib Meas Test
and press
.
3. Use the
softkey to select 200J. Then press the
softkey to charge the defibrillator. Press
.
Readings from defibrillator analyzer:
Read the delivered energy indicated by the defibrillator analyzer. It should
be 0J or be blank.
Readings from M3500B/M5500B display:
The results displayed by the M3500B/M5500B should be as follows:
Available Energy Not recorded - failure if >0.
Msec to charge Ignore - tested earlier
Delivered energy Blank
Impedance Blank
Peak current Blank
Defib errors Not recorded - failure if any reported.
MAIN
ENTER
ENERGY SELECT
CHARGE
DISARM