M3500B/M5500B Service Manual
3
-
11
Error Codes
3
Error Codes
The System Log provides error codes in 5 digit hexadecimal format. Before
replacing any parts, check to see if the cables and flex circuits are properly
connected. See "Servicing Notes" on page 4-1.
NOTE
Confirm any errors by running the self-tests again, making sure that no one
touches the unit during the test.
Table 3-2 Error Codes
Error Code
Meaning
Possible Solutions
00000 - 00400
Defib failure - charging circuits.
1. Replace Power PCA.
2. Replace Control PCA.
01000
Defib failure - biphasic processor.
1. Replace Power PCA.
2. Replace Control PCA.
02000
Leads front end failure.
1. Replace Interface PCA.
2. Replace Control PCA.
04000
Pacer failure.
Replace Power PCA.
08000
Processor error.
Replace Control PCA.
10000 - 1FFFF
System monitor failure.
1. Cycle power.
2. Replace Control PCA.
10001
Processor Synchronization Error
Cycle power.
10004
Synchronization Time Out
Cycle power.
•
If error does not repeat, run ALL
Performance Verification Tests. If
unit passes all tests, the unit can
be returned to service.
•
If error repeats, replace Control
PCA.
20000 - 2FFFF
Front end failure.
1. Replace Control PCA.
2. Replace Interface PCA.
30000 - 3FFFF
Pacer failure.
1. Replace Power PCA.
2. Replace Control PCA.
40000 - 4FFFF
Monitor processor failure.
1. Replace Control PCA.
2. Replace Interface PCA.
50000 - 5FFFF
SpO
2
problem.
1. Replace SpO2 PCA.
2. Replace Interface PCA.
3. Replace Control PCA.
60000 - 6FFFF
Advisory Failure.
Replace Control PCA.
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...