M3500B/M5500B Service Manual
3
-
15
System Messages
3
SpO2 Noisy Signal
Excessive patient movement, electrical
interference, or optical interference is
present.
1. Minimize patient movement or apply the
sensor to a site with less movement.
2. Secure the sensor cable loosely to the
patient.
3. Reduce sources of electrical or optical
interference.
SpO2 Sensor Fail
The SpO
2
cable is not connected to the
device; or the cable or sensor are broken.
1. Attach the cable to the M3500B/
M5500B
.
2. Replace cable and/or sensor.
System Failure - Cycle
Power
A serious error has occurred.
Turn power off, then on. If message reap-
pears, service unit. See "Troubleshoot-
ing" on page 3-1.
Service Unit
Appears during the Shift/System Check
1. Replace Data Card.
2. Replace lithium battery.
3. Replace Control PCA.
Table 3-3 System Messages (Continued)
Message
Description
Corrective Action