Operational Problems
3
-
24
Troubleshooting
Pacing
The following table covers problems that might arise while performing exter-
nal pacing.
Doesn’t deliver correct energy into
Defibrillator Analyzer or delivers no
energy at all. (Should also get mes-
sage -
No shock delivered
or
Defib
failure
.)
Control PCA failure.
Replace Control PCA.
Power PCA failure.
Replace Power PCA.
Doesn’t measure its own delivered
energy correctly.
Power PCA failure.
Replace Power PCA.
Control PCA failure.
Replace Control PCA.
Not synchronizing even though
ECG waveform OK on display.
Control PCA failure.
Replace Control PCA.
Fails
Defib Meas
Test in Diagnostic
Mode (other then symptoms above).
Power PCA failure.
Replace Power PCA.
Control PCA failure.
Replace Control PCA.
Table 3-9 Operational Problems - Defibrillation and Synchronized Cardioversion (Continued)
Message or Symptom
Possible Causes
Suggested solution
Table 3-10 Operational Problems - Pacing
Message or Symptom
Possible Causes
Suggested solution
Doesn’t deliver correct current into
Pacer Tester or delivers no current
at all.
Control PCA failure.
Replace Control PCA.
Power PCA failure.
Replace Power PCA.
Doesn’t measure its own delivered
current correctly.
Power PCA failure.
Replace Power PCA.
Control PCA failure.
Replace Control PCA.
Doesn’t pace at correct rate.
Control PCA failure.
Replace Control PCA.
Pacer hardware not installed.
mes-
sage even though Pacer is present.
Bad connections - Manual Keypad
flex circuits to Interface PCA.
Reconnect flex circuits properly.
See "Servicing Notes" on page 4-1.
Unit fails either the Shift System
Check or Extended Self Test, or
reports error code 90007.
Test sequence is interrupted.
Run the test again, making sure that
no one touches the unit during the
test.
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...