M3500B/M5500B Service Manual
3
-
31
Operational Problems
3
Data Card
The following table covers problems that might arise with the Data Card.
Table 3-17 Operational Problems -Data Card
Symptom
Possible Cause
Corrective Action
Data card is not recognized when
plugged in.
Unit power is already on. Card is
only recognized during power-up
sequence.
Turn power off for 2 minutes, then
on again.
Card is full or not a supported type.
Use a new, empty M3510A/M5510A
Data Card.
Data on the card corrupted.
The card was removed while the
unit power on.
Only remove card after power is
turned off. Delete corrupted files
using the Event Review Data Man-
agement System.
Fails Data Card Test in Diagnostic
Mode
Card is full or not a supported type.
Use a new, empty M3510A/M5510A
Data Card.
Failure on Control PCA.
Replace Control PCA.