M3500B/M5500B Service Manual
4
-
57
Top Case Assembly
4
New parts to be installed - provided with top case
The Replacement Top Case Assembly comes with the following new compo-
nents, which must be installed in the field.
Do not reuse the old parts.
:
z
ECG/SpO
2
connector gaskets (not available separately)
z
ECG/SpO
2
connector retaining clips (M2475-07101)
New parts to be installed - must be ordered separately
In addition, the following parts cannot be reused from the old case, and must
be ordered separately and installed into the new top case:
z
Branding Label Set (includes the Ambient Light sensor cover)
a.
Philips label set (M3500-69556)
b.
Laerdal North America label set (M3500-69557)
c.
Laerdal Other label set (M3500-69558)
z
Manual Keypad
These are available in 2 versions (pacing, no pacing) and in all
the supported languages. See the "Replacement Parts"
chapter,
Table 5-7 and Table 5-8 to select the correct version and
language.
z
Case Label Set
These are available in all the supported languages. See the
"Replacement Parts" chapter,
Table 5-12 to select the correct
language.