Bottom Case Assemblies
4
-
58
Removal and Replacement
Bottom Case Assemblies
The sections that follow describe removing and replacing assemblies that
reside in the Bottom Case.
Battery PCA
The Battery PCA is mounted in the Battery compartment, under the Battery
Cover. It has a cable assembly that enters the case through a rubber gasket.
Once inside the case, the cable assembly connects to the Power PCA and to
the Power Connector.
Removal
1. Unplug 3 connectors from the Battery PCA as shown in
z
4 pin connector to Power PCA (battery connector).
z
2 pin connector to Power PCA.
z
1 in line with wires to Power Connector.
2. Pull the Battery PCA up out of its slot in the bottom case.
3. Pull out the gasket from its hole in the case. This is made easier if the tabs
on the gasket are cut away with a utility knife. Be careful not to gouge the
case.
4. Pull the wires out of the hole in the case.