M3500B/M5500B Service Manual
2
-
5
Test and Inspection Matrix
2
Defibrillator Test -
AC or DC Power
Module (DM)
(Required only
when AC or DC
power module
used.)
Using only the AC or DC
power module, enter Diag-
nostic Mode and run the
Defibrillator Test (at
200Joules) (page 2-27):
If the measurements are as
follows:
Example: DM:p
Displayed by
M3500B/M5500B:
•
Available Energy after
Shock
•
0
•
Msec to charge
•
< 15000 msec
•
Delivered energy
•
Actual delivered energy
200 +7%
•
Impedance
•
42 to 57
Ω
•
Defib errors
•
None (0)
then the Defibrillator test
(Power Module) passes
Defibrillator Test -
Battery Power
(DB)
Using only battery power,
enter Diagnostic Mode and
run the Defibrillator Test (at
200Joules) (page 2-28).
If the measurements are as
follows:
Example: DB:p
Displayed by
M3500B/M5500B
•
Available Energy after
Shock
•
0
•
Msec to charge
•
< 3000 msec
•
Delivered energy
•
Actual delivered energy
200 +7%
•
Impedance
•
42 to 57
Ω
•
Defib errors
•
None (0)
then the Defibrillator test
(Battery Power) passes
Defibrillator
Disarm Test (D)
Enter Diagnostic Mode and
run the Defibrillator Disarm
Test (page 2-29)
If all readings as expected
Available energy - failure >0
No errors reported
then the Defibrillator Disarm
test passes
Example: D:p
Table 2-1 Performance Verification and Safety Tests (Continued)
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
P (pass) or F (fail)
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...