Momentary Messages
3
-
16
Troubleshooting
Momentary Messages
Momentary messages are temporary and only appear on the display for a few
seconds. They are intended for the end user and appear in the
Instructions
for Use
. They are duplicated here for the reference of the service person.
Before replacing any parts, check to see if the cables and flex circuits are
properly connected. See "Servicing Notes" on page 4-1.
Table 3-4 Momentary Messages
Message
Possible Cause
Corrective Action
Attach Leads
An attempt was made to begin pacing in
Demand Mode without ECG leads
attached to the patient.
Attach leads to patient
Attach Pads
The multifunction defib electrode pads are
not making proper contact with the patient.
1. Check the pads are applied to the
patient, as directed on the package.
2. Replace pads if the prompt continues.
Defib Disarmed
The pads connection is compromised.
Check the pads are applied to the patient
properly.
The mode is changed from Manual to
AED while the defibrillator is charged.
If a shock is indicated, deliver the shock
before changing modes.
is not pressed within 30 seconds
of the defibrillator being charged.
To deliver a shock, press
within
30 seconds of the defibrillator being
charged.
is pressed.
None.
No Shock Delivered
Patient impedance is too high or too low.
1. Make sure pads are applied properly.
2. Replace the pads.
3. Replace the pads cable.
Check Printer
Printer paper is absent or jammed; the
printer door is not closed properly.
1. Reload printer paper.
2. Make sure the door is closed properly.
Data Card Full
The incident is more than 2 hours in dura-
tion, causing the Data Card to fill.
None. A new Data Card can not be
inserted during an incident.
An empty Data Card was not inserted for
the incident; the Data Card filled sooner.
Use one empty Data Card per incident/
patient to decrease the chance of the card
filling.
Data Card Interrupted
No data is being recorded on the Data
Card because the Data Card was
removed during an incident.
•
During incident, none. The Data Card
cannot be inserted during an incident.
•
After incident, remove card, turn
power off, insert card, turn power on.
Data Card Not In Ser-
vice
The Data Card is inserted while the
M3500B/M5500B is on.
None. A Data Card must be inserted prior
to turning the M3500B/M5500B on for the
current patient.
Incompatible Data
Card
A Data Card other than the M3510A is
inserted.
Use only M3510A/M5510A Data Cards.
SHOCK
SHOCK
DISARM
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...